- classdef ChannelClass
- % CHANNEL is the class for one channel of the oscilloscope. Multiple
- % settings for the different channels of the oscilloscope can be changed
- % via this class definition
-
- properties (Hidden, Access = private)
- scope % Contains the class of the scope that is parent of this channel.
- channelnumber % is the number of the channel on the scope.
- end
- properties (Dependent)
- state %is channel enabled or not
- coupling %DC or AC
- bandwidth %inputfilter bandwidth
- scale %Voltage scale in volt/div
- offset %DC offset
- probe %probesetting
- label %name for channel on scope
- type %set decimation mode
- probeunit %Define current or voltage probe
- end
-
- methods
- %Constructor of class
- function ch = Channel(scope,channelnumber)
- ch.channelnumber = channelnumber;
- ch.scope = scope;
- end
-
- %% Get.function section
- %all get functions pull information from equipment to update
- %properties.
- function s = get.state(ch)
- s = ch.CHAN('state?');
- end
-
- function out = get.coupling(ch)
- out = ch.CHAN('coup?');
- end
-
- function out = get.bandwidth(ch)
- out = ch.CHAN('band?');
- end
-
- function out = get.offset(ch)
- out = ch.CHAN('offs?');
- end
-
- function out = get.scale(ch)
- out = ch.CHAN('scal?');
- end
-
- function out = get.label(ch)
- out = ch.CHAN('lab?');
- end
-
- function out = get.type(ch)
- out = ch.CHAN('type?');
- end
-
- function out = get.probeunit(ch)
- out = ch.scope.query(['PROB',num2str(ch.channelnumber),':SET:ATT:UNIT?']);
- end
-
- function out = get.probe(ch)
- out = ch.scope.query(['PROB',num2str(ch.channelnumber),':SET:ATT:MAN?']);
- end
-
- %% Set.function section
- % This section contains all functions to change settings of a
- % channel
-
-
- function ch = set.state(ch,in)
- %STATE Set the channel on or off.
- %Options: ON | OFF
- ch.CHAN('state',in);
- end
-
-
- function ch = set.coupling(ch,in)
- %COUPLING Set coupling to one of the following settings:
- %Options: DCLimit | ACLimit | GND
- ch.CHAN('coup',in);
- end
-
-
- function ch = set.bandwidth(ch,in)
- %BANDWIDTH Set bandwidthlimit. Enable lowpass 20MHz filter
- %Options: FULL|B20
- ch.CHAN('band',in);
- end
-
- function ch = set.offset(ch,in)
- %OFFSET Sets the DC offset for channel
- %Range: Depend on vertical scale and probe attenuation.
- %Increment: Depends on vertical scale and probe attenuation.
- %On reset: 0
- %Default unit: V
- ch.CHAN('offs',in);
- end
-
- function ch = set.scale(ch,in)
- %SCALE Sets the scale for the channel
- % Scale value, given in Volts per division.
- % Range: 1e-3 to 10 (without probe attenuation)
- % *RST: 5e-3
- % Default unit: V/div
- ch.CHAN('scal',in);
- end
-
- function ch = set.label(ch,in)
- %LABEL Sets label of the channel on the scope
- ch.CHAN('lab',in);
- end
-
- function ch = set.type(ch,in)
- %Sets the type of the channel
- ch.CHAN('type',in);
- end
-
- function ch = set.probeunit(ch,unit)
- ch.scope.write(['PROB',num2str(ch.channelnumber),':SET:ATT:UNIT ', unit]);
- end
-
- function ch = set.probe(ch,man)
- ch.scope.write(['PROB',num2str(ch.channelnumber),':SET:ATT:MAN ', man]);
- end
-
- %% not set or get function
-
- function enable(ch)
- %ENABLE channel
- ch.state('ON');
- end
-
- function disable(ch)
- %DISABLE channel
- ch.state('OFF');
- end
-
- function out = frequency(ch)
- %FREQUENCY gets the measured frequency of current signal.
- %Equal to one over the period.
- % See also PERIOD
- out = str2double(ch.MEAS('freq'));
- end
-
- function out = peak2peak(ch)
- %PEAK2PEAK gets the peak2peak value of current signal.
- %Equal to AMPLITUDE times two.
- out = str2double(ch.MEAS('peak'));
- end
-
- function out = period(ch)
- %PERIOD Gets the period of the current signal.
- %Equal to one over the frequency.
- out = str2double(ch.MEAS('per'));
- end
-
- function out = amplitude(ch)
- %AMPLITUDE Gets the amplitude of the current signal.
- %Equal to half the PEAK2PEAK value.
- %See also PEAK2PEAK
- out = str2double(ch.MEAS('ampl'));
- end
-
- function out = mean(ch)
- %MEAN Gets the mean value of the current signal.
- out = str2double(ch.MEAS('mean'));
- end
-
- function out = rms(ch)
- %RMS Gets the rms value of the current signal.
- out = str2double(ch.MEAS('rms'));
- end
-
- function out = phase(ch)
- %PHASE gets the phase shift between two signals.
- %Cannot be done on a single channel.
- out = str2double(ch.MEAS('phas'));
- end
-
- function data = waveform(ch,window)
- %WAVEFORM allows for the waveform to be downloaded to matlab
- % data = WAVEFORM(ch,window) gets the waveform from the scope and
- % returns a array of the waveform.
- %
- % WINDOW can be one of the following options:
- %
- %*DEFault* (Default option)
- % Waveform points that are visible on the screen. At maximum
- % waveform rate, the instrument stores more samples than visible
- % on the screen, and DEF returns less values than acquired.
- %
- %*MAXimum*
- % All waveform samples that are stored in the memory. Only available
- % if acquisition is stopped.
- %
- %*DMAXimum*
- % Display maximum: Waveform samples stored in the current
- % waveform record but only for the displayed time range. At maximum
- % waveform rate, the instrument stores more samples than
- % visible on the screen, and DMAX returns more values than DEF.
- % Only available if acquisition is stopped.
- %
- % See also OSCILLOSCOPE.SETWAVEFORMSETTINGS
- ch.scope.single;
- if nargin < 2
- window = 'DEF';
- end
- ch.scope.setWaveformSettings(window,'REAL')
- data = ch.getWaveform;
- end
-
-
-
- end
-
- methods (Hidden, Access = private)
- function c = CHAN(ch,string,in)
- % CHAN checks if the string corresponds with a query or write
- % c = CHAN(ch, string, in) Where 'ch' is the channel-object,
- % 'string' is a part of the SCPI-command and 'in' is the actual
- % value is to be set. 'in' is optional.
-
- if nargin == 2 %If only ch and string are given it can be a query
- if strcmp(string(end),'?') %If the string ends with a '?' then it is a query.
- c = ch.scope.query(['CHAN',num2str(ch.channelnumber),':',string]);
- else %Without 'in' and no string it is a event.
- ch.scope.write(['CHAN',num2str(ch.channelnumber),':',string]);
- end
- else %If there is a third argument then the setting should be uploaded to the scope.
- ch.scope.write(['CHAN',num2str(ch.channelnumber),':',string,' ',in]);
- end
- end
- end
-
-
- methods (Hidden)
-
- function data = getWaveform(ch)
- %GETWAVEFORM Downloads and processes the waveform from scope.
- header = str2double(ch.CHAN('DATA:HEAD?')); %downloads the dataheader of the waveform that is stored on the scope.
- data.start = header(1); %First position in header is starttime in seconds
- data.stop = header(2); %Second position in header is stoptime in seconds
- data.length = header(3); %Thirth position in header is record length of the waveform in samples.
- data.sampletime = str2double(ch.CHAN('DATA:XINC?')); %Sample time is downloaded from scope.
- ch.scope.opc; %Wait for scope.
- ch.scope.write_unsafe(['CHAN',num2str(ch.channelnumber),':DATA?']); %Send command to download the data.
- %NOTE! Must be done with write_unsafe. Normal WRITE would also
- %check if there are errors but this will interfere with the
- %datastream.
- data.data = ch.scope.readWaveform(data.length); %Read data from buffer.
- end
-
- function c = MEAS(ch,string)
- %MEAS makes all the measurement-functions easier.
- %Most of the measurements queries start with
- %MEAS<m>:RESult:ACT? Thus goes automatic right now.
- c = ch.scope.query(['MEAS',num2str(ch.channelnumber),':RES:ACT?',string]);
- end
-
- % function c = CHANsend(ch,string,in)
- % c = ['CHAN',num2str(ch.channelnumber),':',string,' ',in];
- % end
- end
- end
|